 |
 |
 |
 |

Comparisons Between
Ray-tracing and Reality
At Akustikon
we are continously following up the reliability of our room acoustical
computer programmes. We have compiled 67 comparisons between predicted
and measured echograms in 8 different halls. This work, presented at
the International Congress on Acoustics 1995, shows that the
ray-tracing technique is a reliable design tool and that most of the
room acoustical parameters reveal very good agreement.
The calculation
input data were not modified using the measurement results.
Consequently, the results are not only a pure test of the programme but
also a test of the skill of the operator. The results presented are
therefore not directly comparable to other tests where input data has
been optimised based on measurements in the real hall.
Ray-tracing
Since 1984 we have developed and used our own ray-tracing programme.
The main advantage of our programme compared with commercial available
programmes is the facility to simulate almost any type of surface. The
programme contains eight different surface properties which also can be
combined on the same surface. Even true concave focusing planes are
included. For more information see Technical
Note No. 1.
Ray hits on audience area. Concentration in the rear part of the hall
in the time interval 50-100 ms after the direct sound. Gothenburg
Concert Hall (built in 1935).
Measurements
Most of the measurements of the impulse response were performed with
the MLSSA system which uses a pseudo-noise sequence of the maximum
length type (MLS). The sound source was a dodecahedron loudspeaker with
12 four inch elements. This source is essentially omnidirectional up to
the 2kHz octave band. The measurements were BP-filtered with a three
octaves broad band centered at 1kHz (354Hz - 2828Hz). Note that in some
cases the parameter values may differ considerably for the octave bands
500Hz, 1kHz and 2kHz.

Measurement in the
Concert hall at Artisten, Gothenburg, Sweden.
Parameter Comparisons
Figure 1 shows the relative error for the calculated parameters. A
Relative error less than one means that the absolute difference between
the calculated values and the measured values is less than the audible
difference. The parameters are Definition (D) - speech intelligibility,
Clarity (C) - clarity for music, Lateral Fraction (LF) - spaciousness
and apparent source width, Centre Time (Ts) - an early to late energy
measure, Early Decay Time - percieved reverberation and Strength (G) -
relative sound level.

Figure 1. Error of the calculated parameters related to the subjective
difference limen.
The values of the estimated room acoustical parameters using
ray-tracing and the corresponding measured data are plotted in scatter
diagrams in figure 2.

Figure 2. Scatter diagram of the room acoustical parameters. The blue
line represents perfect agreement. The red line indicates the least
square fit to the data points (linear regression).
The correlation coefficients given in the diagrams indicate the degree
of linear dependence. However, high correlation is not sufficient for
good agreement. On the other hand low correlation does not necessarily
imply useless values. This is illustrated by the parameter LF which has
a fairly low correlation coefficient of 0.65 but the mean absolute
difference is comparable to the subjective difference limen (see Table
1 below).
|
D
|
C
|
LF
|
Ts
|
EDT
|
G
|
| Number of comparisons |
67
|
67
|
50
|
67
|
54
|
25
|
| Correlation coefficient |
0.912
|
0.928
|
0.651
|
0.959
|
0.906
|
0.995
|
| Mean absolute difference |
0.047
|
0.74 dB
|
0.06
|
6.7 ms
|
22.1%
|
0.55 dB
|
| Std dev of the differences |
0.040
|
0.66 dB
|
0.05
|
6.1 ms
|
14.5%
|
0.46 dB
|
| Subjective difference
limen |
0.05
|
0.5 dB
|
0.05
|
10 ms
|
5%
|
1 dB
|
Table 1. Results of the comparisons and order of magnitude of
subjective difference limens.
Echogram Comparisons
In figure 3 below some calculated and measured echograms are plotted.
The diagrams clearly show that even the fine structures of the
echograms are very well matched. The calculations are smoothed by
averaging within fixed 5 ms time windows. The measurements are
BP-filtered with a three octaves broad band centered at 1kHz (354Hz -
2828Hz) and smoothed with a 5 ms moving average window.

Figure 3. Comparisons between calculated and measured echograms. Top
left: Tonhallen, Sundsvall, Sweden
Top right: The audiotorium at the Univ. of Uppsala, Sweden
Bottom left: The Civic hall of Västervik, Sweden
Bottom right: The Semper Opera, Dresden, Germany.

|
|